Honeywell Vertex M User Manual page 150

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

Event
Event List Desc.
Event History
Number
(Short)
Description
INVALID SUP
Supply Pressure
226
PRES
Out Of Range
NO PYRO
No Pyrolyzer
228
MAINS PWR
Mains Power
OPTICS CFG
Invalid Optics
229
ERR
Configuration
No
COM FAIL TO
230
Communication
PLC
From PLC To Az
AUTO LED
231
DRV ADJ FAIL
Adjustment Failed
Q-Factor Out Of
232
BAD Q-FACTOR
Range
FAIL OPT
Optics Verification
233
VERIFY
Failed
Light Gray Read
234
GRAY FAILED
Failed
Data Field
Problem
in. Hg
Manifold pressure out of range
1
No mains power at points 5-8
2
No mains power at points 1-4
3
No mains power at all 8 points
Microprocessor in optics block
Error code
reports problem.
0.0
No communication from PLC
Optics counts
LED adjustment failed
Q-Factor
Q-Factor out of range
Adj Cnts
Optics Verification Failed
Optics values of this field were
Optics Cnts
outside limits
Possible Cause
Miscalibrated sensor board or defective
Replace Analyzer
transducer
Internal thermal fuse opened
Check cooling fan, replace thermal fuse
Pyro mains switch not on
Turn on pyro mains switch
Circuit breaker tripped
Breaker is self resetting
Power connector disconnected
Check connection on back of Analyzer
Defective transformer
Replace transformer
Reread Chemleader.
Chemleader read failed
Cycle power to analyzer.
Update software in optics block.
Cycle power to the analyzers and reload Configuration Profile
Communication has timed out to PLC
Check FLT light on PLC
Connection at rear of analyzer
Check 25-pin, multifunction cable at the rear of the analyzer
Check cable connections to the PLC
Check cable connection at chassis from carrier
Tape incorrectly loaded
Reload and retry tape load
Dirty optics block
Service or replace analyzer
Bad optics PCB set
Service or replace analyzer
Bad RFID tag
Load new Chemcassette
Chemcassette leader not tight or
improperly positioned during white to
Reload Chemcassette
light gray calibration
Bad RFID tag
Load new Chemcassette
Dirty optics block
Clean and recalibrate
Bad optics PCB set
Service or replace analyzer
Chemcassette leader not tight or
improperly positioned during light gray to
Reload Chemcassette
dark gray calibration
Bad RFID tag
Load new Chemcassette
Dirty optics block
Clean and recalibrate
Bad optics PCB set
Service or replace analyzer
Chemcassette leader not tight or
improperly positioned during light gray to
Reload Chemcassette
dark gray calibration
Bad RFID tag
Load new Chemcassette
Dirty optics block
Clean and recalibrate
Bad optics PCB set
Service or replace analyzer
Recovery