Honeywell Vertex M User Manual page 144

Continuous gas monitor
Hide thumbs Also See for Vertex M:
Table of Contents

Advertisement

Event
Event List Desc.
Event History
Number
(Short)
Description
Timeout Run-
TIMEOUT PNT
104
time Point
DIS
Disable
Timeout Run-
TIMEOUT ALM
105
time Alarm
DIS
Disable
POS PNT
Positive Point
106
PRESSURE
Pressure
FLOW ADJ ERR
High Flow After
108
HI
Adjustment
Chemcassette
109
CC EXPIRED
Passed Expiration
Date
Invalid
110
INVALID CC
Chemcassette
No
111
DACS COM FAIL
Communication
from PC To Az
Single pump
SWAPPED BAD
112
failed and
PUMP
swapped
Data Field
Problem
0.0
Operator error
0.0
Operator error
Positive pressure between
sampling location and monitor
Pressure in. Hg
Transducer error
Offset incorrect
Flow cannot be adjusted below
cc/min.
210 cc/min.
Operator error
0.0
Chemcassette expired
CC Tag does not match profile
or is unreadable
0.0
Ethernet communications
Pump swap has occurred
Number of bad pump -
because vacuum levels fell
1 or 2
below 6 in. Hg
Possible Cause
Time limit too short
Point disabled longer than time limit setting in the Configuration Profile
Time limit too short
Point disabled longer than time limit setting in the Configuration Profile
Time limit too short
Point pressure above atmospheric pressure while in idle
Calibration error
Defective sensor
Debris in proportional valve
Debris or defective valve Proportional valve held open
Chemcassette installed past its expiration date
Expiration date reached
Bad configuration or corrupted RFID
Communications to DACS interrupted longer than 20 seconds
Pump failure
Unused slots not plugged
Large 0.018 in transport flow orifices installed
Recovery
Change time limit in
Configuration Profile
Press RESET to reinitialize
timer
Restore point to Monitor Mode
Change time limit in
Configuration Profile
Press RESET to reinitialize
timer
Restore point to Monitor Mode
Change time limit in
Configuration Profile
Relieve/reduce pressure
Contact Honeywell Service
Contact Honeywell Service
Contact Honeywell Service
Verify range of valve
Perform Auto Balance to clear
debris
Replace with 0235-1279
Replace Chemcassette
Replace Chemcassette
Check gas configuration
Replace Chemcassette
Check Ethernet cable at rear of
analyzer
Check Ethernet hub
connection and operation
Check Ethernet connection to
DACS
Restart DACS to clear OPC
driver problem
Rebuild/Replace non-
operating pump
Install 1295-0404 connector
plug on unused slot
Replace all orifices with 0.015
in orifices (p/n 1290K0009 per