Preparatory measures before (initial) start-up

5.3.2 Removing the battery from the device

Only change the power module when the device is switched off and the
Note
defibrillation electrode plug is disconnected.
Fig. 9: Removing the battery
Procedure:
Lay the device on its back and press the release button (as described in 5.3.1) to
the right until the tongue of the power module is released and the power module
snaps out of the slot.
Twist the power module slightly in the direction of the arrow (1st) and then pull it
in the direction of the arrow (2nd) out of the device.
Operating instructions
HeartSave PAD 22289 / GB / C
CAUTION
Monitor the Status Display. If the display is showing "OK",
the device is ready to use.
(If applicable), switch it off using the on / off button or put
the device cover back on.
If the display does not show "OK", remedy the cause or
contact your nearest service station.
The device will switch itself off.
25