Preparatory measures before (initial) start-up
5.4 PRIMEDIC™ Battery
The battery is a disposable lithium battery. It is fully charged when delivered. This type of battery
is state-of-the-art and was selected due to its extremely long service life and power storage.
In any event, heed the instruction leaflet enclosed with the battery and keep it safe
with these operating instructions.
Note
Operating instructions
26
On no account try to charge the battery.
There would be risk of an explosion!
Use the battery before its use-by date expires.
After using the device, the battery should be (if necessary)
exchanged for a new one (to guarantee a full period of
operation when next deployed).
If the device has to be sent away to technical services, remove the battery
before sending it and put some adhesive insulation tape over its contacts.
When sending the battery, observe the separate shipping regulations.