Connecting; General Connection Information - Eizo RadiForce LS560W Instruction Manual

8mp 56" color flat panel monitor
Table of Contents

Advertisement

Connecting

6.1

General connection information

All information and warnings related to this product must be observed to ensure danger-
free operation.
CAUTION
Changes to device
• Device settings may only be adjusted by trained service personnel; otherwise, the
• Do not make any changes to the device without prior approval from the manufacturer.
CAUTION
Observe shielding measures
The shielding measures described in the applicable national EMC guideline must be
observed. If these guidelines are not observed, malfunction of the monitor may result.
Observe the grounding measures
To ensure that the permissible leakage current is not exceeded in the case of a fault, you
must additionally ground the monitor with a separate ground line.
Information on cable installation
• Only shielded cables are permitted for all signal connections.
• If the relevant facility is available on the connector, all plug connections must be
• If signal and power supply cables are routed next to one another, reversible pixel errors
• The display must not share a power supply with motors or valves (interference!).
• Externally connected cables can represent a trip hazard. Make sure that all incoming
• Cable grips are provided in the device. Use these to secure the cables against
RadiForce® LS560W
Instruction Manual, 04/2012, A5E10000920A-006
CAUTION
warranty is void.
screwed tight or locked.
may occur if there is high interference on the power supply network.
cables are safely routed.
unintentional loosening.
6
23