Page 1
Instruction for Use 8MP 58" LCD Monitor Important Please read the Safety Information and the information delivered with the product carefully to familiarize yourself with safe and effective usage.
Note the following: WARNING EIZO products may only be used for the applications described in the catalog and in the relevant tech- nical documentation. If products and components from other manufacturers are used, these must be recommended or approved by EIZO. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
The monitor should be mounted in a ceiling suspension unit or wall mount. The RadiForce LS580W is intended to be used by health care professionals to display the video sources from various commercially-available devices (with the exception of mammog- raphy) commonly used in a medical environment on a single video monitor.
2.1 General safety notes Correct transport, professional storage, installation and connection, as well as careful opera- tion and maintenance, are required to ensure that the EIZO devices operate safely and cor- rectly. The devices may only be used for applications for which they are commonly used.
Page 7
Touching the components may be fatal. No contact between device and patients The device is not suitable for direct contact with patients. Device and patient must never be touched simultaneously. Otherwise there is a danger to life and limb. RadiForce LS580W Instruction for Use, 02/2014...
Page 8
Only use power cables approved for China. These power cables are identified by the labels "CCC" or "CQC". Observe the country-specific regulations Observe all regulations of the country in which the device is used. RadiForce LS580W Instruction for Use, 02/2014...
Page 9
If the device is brought into a room with a higher or rising temperature, condensed water will form in and on the device. Do not switch on the device until the condensed water has evaporated. Otherwise the device could be damaged. RadiForce LS580W Instruction for Use, 02/2014...
Page 10
Note that monitors can fail as a result of aging, and that image properties such as brightness, con- trast, and color value can change. Do not touch the monitor screen Due to mechanical pressure or electrostatic discharging, touching the screen can result in brief dis- turbances to the image. RadiForce LS580W Instruction for Use, 02/2014...
This is a class A device. The device may cause radio interference or interfere with the operation of other devices in close proximity. In this case the user is encouraged to perform appropriate measures to correct the interference. RadiForce LS580W Instruction for Use, 02/2014...
Note Keep the packaging material for subsequent transport of the device. Device The RadiForce LS580W is a LCD Monitor for mounting in a ceiling suspension unit or wall mount. Product Order No.
Page 13
Graphics settings are also retained, for example after changing the graphics card or PC, and the monitor does not have to be calibrated. RadiForce LS580W Instruction for Use, 02/2014...
3.3 Medical subsystem NOTICE Only trained and qualified personnel The device may only be used by trained and qualified personnel. The RadiForce LS580W can be part of a medical subsystem comprising the following com- ponents. Mandatory Devices • RadiForce LS580W •...
CAUTION Changes to device Do not make any mechanical or electric changes to the device. EIZO GmbH will not be held liable if changes are made to the device. 4.1 Installation site Note the following conditions at the installation location. NOTICE...
• Make sure the monitor cannot tip over. • The monitor must be removed from the packaging and carried by at least two persons. • Wear appropriate protection to prevent injuries should the monitor fall. RadiForce LS580W Instruction for Use, 02/2014...
Page 17
1. Carefully open the packaging and remove the upper section of the carton as well as all accessible packing parts. 2. Lift the monitor from the base of the packing carton. Hold it on the side and underside. RadiForce LS580W Instruction for Use, 02/2014...
• An installed stand must be sufficiently stable such that tilting up to 10° does not result in the monitor toppling. Note Recommended wall mount EIZO recommends the FWM6300 wall mount because it has been tested with the monitor and meets all necessary conditions. Order No.: 6GF6988-8AB02 Note...
CAUTION Changes to device Do not make any mechanical or electric changes to the device. EIZO GmbH will not be held liable if changes are made to the device. CAUTION Shielding measures Follow all shielding measures according to the country-specific EMC Directive. If these guidelines are not observed, device malfunction may result.
Page 20
• Externally connected cables can represent a trip hazard. Make sure that all incoming cables are safely routed. • If the device includes cable grips, please use them to prevent unintended loosening of connected cables. See also Safety instructions [} 6] Electromagnetic compatibility [} 35] RadiForce LS580W Instruction for Use, 02/2014...
The device's power supply is connected using an appliance plug. DVI connectors The monitor has two dual-link DVI connectors (A and B) 5 V connections The monitor has two 5 V outputs which can be used to supply external devices. RadiForce LS580W Instruction for Use, 02/2014...
2. Remove the connector panel cover. 3. Connect the appliance plug to the monitor power socket. 4. Secure appliance cable with cable ties to the lug to prevent unintentional loosening (marked with a red circle). RadiForce LS580W Instruction for Use, 02/2014...
Page 23
5. Insert the DVI cable directly into the DVI socket. Only Dual Link DVI cables of a corre- spondingly high quality may be used. 6. If you use TDL3600-QL modules from EIZO: Connect the CAT cables to the DVI receiver module connectors and tighten the clamps.
If the operation LED does not light green after the equipment has been switched on and a video signal has been applied, check the system for basic connection and operating errors before contacting service per- sonnel. RadiForce LS580W Instruction for Use, 02/2014...
EDID data (Extended Display Identification Data) of the monitor is transferred to the graphics card. You can now configure the resolution based on the driver or operating system settings. RadiForce LS580W Instruction for Use, 02/2014...
6.4 Check for pixel defects Pixel defects (small bright or dark dots) can occur in LCD monitors. During the manufactur- ing process, all monitors are checked for the permitted number of defective pixels. Defective pixels cannot be corrected. RadiForce LS580W Instruction for Use, 02/2014...
Measures in the event of a failure Note Device malfunction in operation If the device is not working properly, check the system for incorrect connection or opera- tion before contacting Service. RadiForce LS580W Instruction for Use, 02/2014...
Quartamon Med 0.5% Chlorhexidin in 70% Iso- propylalcohol Disinfecting agents Perform 3% by weight Morning Mist (1:64) Terralin Protect 2% by vol. Mikrozid Sensitive Liquid, undi- luted Microbac Tissues Glucoprotamin Incidin Plus 8% by vol. RadiForce LS580W Instruction for Use, 02/2014...
Page 29
Further examples tion agents: Light gasoline Cleaning solvent, Boiling range Petroleum ether gasoline Note Cleaning other components Information on cleaning or disinfection of other system components can be obtained from the respective instructions for use. RadiForce LS580W Instruction for Use, 02/2014...
Correct the settings if necessary. Carrying out quality tests All quality tests can be performed using the EIZO RadiCS program. Checking, changing, and calibrating settings You can check and change the monitor settings and calibrate them if necessary using a photometer and suitable software.
Lamp warm-up period: Inform servicing department Setting is active, and the warm-up pe- riod has expired without the monitor having reached the stabilized lumi- nance level. Internal error Inform servicing department RadiForce LS580W Instruction for Use, 02/2014...
AC 100 ... 240 V (± 10 %) Line frequency 50 ... 60 Hz (± 5 %) Current consumption < 2.0 A @ 240 V / 5.0 A @ 100 V Maximum current consumption 400 W Energy saving mode < 41 W RadiForce LS580W Instruction for Use, 02/2014...
Dimensions (W x H x D) in mm 1319 x 776 x 146 10.6 Climatic conditions In operation Temperature range +5 °C ... +40 °C Ambient temperature Temperature gradient Max. 5 °C/h, no condensation Air pressure 570 ... 1050 hPa RadiForce LS580W Instruction for Use, 02/2014...
IEC/EN 60601-1 (Third Edition) • CAN/CSA - C 22.2 No. 60601-1-08 • GB4943.1 (non-tropical, altitude < 2000 m) Protection class Protection class I Degree of protection IP20 Conformity CE according to MDD 93/42/EEC (Class I) RadiForce LS580W Instruction for Use, 02/2014...
EN 61000-4-11 Line reaction to harmonics EN 61000-3-2:2006 + A1:2009 + A2:2009 GB17625.1 Line reaction to voltage fluctuations EN 61000-3-3 Limits of radio disturbance characteristics GB9254 See also Guidance and manufacturer's declaration – electromagnetic emissions [} 38] RadiForce LS580W Instruction for Use, 02/2014...
Dimensional drawings 11.1 Dimensions of RadiForce® LS580W 11 Dimensional drawings All dimensions in mm 11.1 Dimensions of RadiForce® LS580W Front view RadiForce LS580W Instruction for Use, 02/2014...
Page 37
Dimensional drawings 11.1 Dimensions of RadiForce® LS580W View from behind, below, and to the side - with cover RadiForce LS580W Instruction for Use, 02/2014...
12 Appendix 12.1 Guidance and manufacturer's declaration – electromagnetic emissions The RadiForce LS580W monitor is intended for use in the environment specified below. The user of the monitor should ensure that it is used in such an environment. Emissions test Compliance Electromagnetic environment –...
Marking according to ACPEIP (China-RoHS). (name plate) "On" symbol (voltage) (power switch) "Off" symbol (voltage) (power switch) Input for service calls. (PS2 socket) Symbol "Comply with the instruction manual". (device) Symbol for "Authorized service personnel only". (device) RadiForce LS580W Instruction for Use, 02/2014...
Please observe all local requirements and laws pertaining to the disposal of devices. 12.6 Accessory devices Connected devices such as PCs must meet the relevant safety standards. 12.7 Contact Support during installation and for technical questions Medical Solutions (http://www.eizo.com) RadiForce LS580W Instruction for Use, 02/2014...
12.8 Trademarks 12.8 Trademarks EIZO Logo is a registered trademark of EIZO Corporation in Japan and other countries. EIZO is a registered trademark of EIZO Corporation in Japan and other countries. RadiForce is a registered trademark of EIZO Corporation in Japan and other countries.
EFUP shall be done in accordance with the Product Mainte- nance Procedures. This product must not be disposed of as unsorted municipal waste, and must be collected separately and handled properly after decommissioning. RadiForce LS580W Instruction for Use, 02/2014...
Page 43
(economically or technically) available sub- stitutes. 产品中有毒有害物质或元素的名称及含量 Table of hazardous substances’ name and concentration. RadiForce LS580W Instruction for Use, 02/2014...
Need help?
Do you have a question about the RadiForce LS580W and is the answer not in the manual?
Questions and answers