Application Planning; Installation Site - Eizo RadiForce LS560W Instruction Manual

8mp 56" color flat panel monitor
Table of Contents

Advertisement

Application planning

4.1

Installation site

Provide adequate ventilation
Ventilation holes are located on the rear of the housing.
If the ventilation holes are covered or closed, the heat generated in the device cannot be
dissipated sufficiently.
● Do not cover the ventilation holes.
● Do not close the ventilation holes.
Ensure unimpeded access to the power switch
CAUTION
Power switches and connections must be accessible at all times
When assembling or installing the device, ensure that the power switch and the
connections are accessible at all times.
Avoid dusty environments
The monitor has been designed for use in the clean environment of medical diagnostics. The
display dissipates heat through the openings at the rear. Dust from dirty environments can
penetrate into the monitor through these openings. In the extreme case, deposits are
possible which become evident as dark spots in a white picture and which can result in
deterioration of the luminance.
● Protect the display from dust, e.g. during building measures at the installation location.
Use the original packaging or service packaging for transport.
Observe the permissible ambient temperature range
The permissible ambient temperature range must not be violated.
RadiForce® LS560W