Application planning

4.1

Installation site

Ensure unimpeded access to the power switch
CAUTION
Power switches and connections must be accessible at all times
When assembling or installing the device, ensure that the power switch and the
connections are accessible at all times.
Provide adequate ventilation
Ventilation holes are located on the rear of the housing.
In order to guarantee adequate air circulation, the back of the device must be at least 10 cm
away from the wall or 15 cm away from another device.
Avoid dusty environments
The display has been designed for use in the clean environment of medical diagnostics. The
display dissipates heat through the openings at the rear and top. Dust from dirty
environments can penetrate into the display through these openings. In the extreme case,
deposits are possible which become evident as dark spots in a white picture and which can
result in deterioration of the luminance. Protect the display from dust, e.g. during building
measures at the installation location, and use the original packaging for transport.
Observe the permissible ambient temperature range
The permissible ambient temperature range must not be violated.
Avoid reflections on the screen
The display has an anti-glare surface. The anti-glare surface is only effective if the screen is
clean and grease-free.
The display should be positioned such that reflections of lights, windows, furniture with shiny
surfaces or light-colored walls do not appear on the screen.
In order to reduce mirroring on the unit, only use non-dazzling reflector lights for the ceiling
lighting.
GX1030
Instruction Manual, 11/2011, 1020008-002
4
17