Battery Monitoring And Charging; Status Indication Of The Device; A Ünzeige Des Berwachung Und Gerätestatus Wiederaufladen Der; Atterie Leistungsfähigkeit Der Nzeige Des Gerätestatterietus Batterietterie - IGEA I-ONE User Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
IGEA
5.3

Battery monitoring and charging

The I-ONE can also charge the battery during treatment.
In the top of the display, the battery symbol is always shown indicating the percentage
of remaining charge. The symbol is normally green and turns red when the battery
needs charging.
If the external charger is connected during treatment, a flash appears inside the battery
symbol to indicate that charging is in progress, and there is a beep as a warning that the
power supply is being switched on.
With the generator switched off, to charge the battery, connect the power supply first
to the generator and then to the mains socket.
until the battery charge is complete.
As it is normal for the battery to heat up when charging, if treatment is to be carried out whilst charging, it is
recommended not to place the I-ONE in direct contact with the body.
5.4

Status indication of the device

Display
The display lights up and the generator beeps and
vibrates.
The display shows the animation of the battery being
charged and a flash symbol below it
The animation persists until the charge is complete (full charge may take up
to 3 hours)
When charging is complete, the display shows the battery charge symbol →
Disconnect the power supply unit from the generator and the mains socket.
If the I-ONE is switched on during charging, pressing the power button turns on the
generator and the user can continue with the treatment whilst the battery is being
charged. Under these conditions, the device behaves as described in section 5.2.
At the end of treatment, leave the I-ONE switched off and connected to the power supply
Generator on during treatment
LED and alarms
The LED at the top is ON with a flashing
green light
The timer appears on the display with the
treatment time decreasing every minute
and the white bar rotates inside the
timer.
The PAUSE symbol is visible below the
timer.
Pressing
it
treatment delivery.
Device status
The I-ONE is switched on and is
providing the treatment.
The battery symbol in the top centre
shows the percentage of charge
remaining; it is green up to 20% and
red when the capacity is > 20%.
interrupts
the
I-ONE - User Manual
Page 8
9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ig4160n

Table of Contents