Installation Information - Sartorius PMA.Evolution Installation Instructions Manual

Paint-mixing scales for potentially explosive areas zone 1
Hide thumbs Also See for PMA.Evolution:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 3
The IP protection rating of the scale and
the ex-link converter YCO14-Z is IP40 as per
EN 60529. The device must be handled carefully
according to the IP protection rating.
The environment must be suitably secured.
Ex Zone 1 (Category 2 Equipment)
− In accordance with Directive 94/9/EC, the EVO1X | LAB1X
model is a category 2 device, suitable for use in Zone 1
potentially explosive areas.
EC Type Examination Certificates: FM14ATEX0085X
ID code: II 2G Ex ib IIB T4 Gb
− The ex-link converter is only suitable for installation as an
electrical apparatus with the following ID code outside of
the potentially explosive area:
II (2)G [Ex ib Gb]IIB
as per EC Type Examination Certificate no.
FM14ATEX0084X
The converter may only be connected to supply voltages of
90 V to 264 V at a frequency of 48–62 Hz.
If the device is used in Zone 1 potentially
explosive areas outside the Federal Republic of
Germany, the relevant national electrical codes
and safety regulations must be observed. Ask
your dealer or Sartorius Service Center about the
guidelines that apply in their country.

2.2 Installation Information

Do not operate the scale if its housing, the
ex-link converter, or the AC adapter including
all connections are damaged. Immediately
disconnect the damaged device from the power.
Do not expose the scale, the ex-link converter,
or its AC adapter or the accessories supplied by
Sartorius to extreme temperatures, aggressive
chemical vapors, moisture, shocks, vibrations,
or strong electromagnetic fields. Observe the
conditions of operation described in the
Specifications.
The casing on all connection cables as well as
the casing on the wires inside the equipment
housing are made of PVC. Chemicals that
corrode this material must be kept away from
these cables.
The operator shall be solely responsible for any
modifications to the equipment and for
connecting any cables or equipment not
supplied by Sartorius! Information on
operational quality is available upon request
from Sartorius.
You should only use accessories supplied by
Sartorius.
Note the IP protection rating of the scale, the
ex-link converter, and its AC adapter. Do not
allow liquid penetration. The protection rating
specifies the suitability of equipment for various
environmental conditions (moisture, foreign
bodies).
4
EVO1X | LAB1X Installation Instructions
Before cleaning the AC adapter, ex-link
converter, or the scale:
Disconnect all devices from the power.
The scale and ex-link converter should only be
opened by personnel trained by Sartorius with
the power disconnected. Do not open the AC
adapter.
Avoid generating static electricity on the glass
panel of the touch screen and plastic casing.
The equipotential bonding conductor of the
devices must be connected properly, according
to commonly accepted technical standards.
Only clean the device as stipulated in the
cleaning instructions.
Take care that the glass panel of the touch
screen is not damaged (e.g., by falling objects,
impact, or extreme pressure).
If the glass panel is damaged, disconnect the
device from the power supply immediately.
The surface of the touch screen should not be
touched with pointed, sharp, hard, or rough
objects. You should only use the touch pen
provided or your fingertip. Do not use parts of
clothing (e.g., sleeves) or sponges for cleaning
because these can scratch the surface (e.g., due
to rivets, buttons on the sleeve, or sand in the
sponge).
The device must be protected from unnecessarily
extreme temperatures, aggressive chemical
vapors, moisture, shocks, and vibrations. Note
the connection data (see EC Type Examination
Certificates for the device and/or the safety
instructions, drawing no. 2023040).
Warnings Concerning Installation and Operation:
The equipment must only be used indoors. Avoid
generating static electricity on glass and plastic
parts. Connect the scale and the ex-link
converter (YCO14-Z) to the equipotential
bonding conductor using a suitable low
resistance method. All electrical circuits are
grounded and electrically connected to the
metal parts of the device.
− The installation must be inspected for
correct function and safety by a trained
and qualified person at appropriate intervals
(e.g., checking the cable for damage).
− Operating personnel must be trained to
recognize faulty operating states and to be
able to initiate the necessary safety measures
(e.g., disconnecting the ex-link converter
from the power supply).