Maintenance
October 2023
membrane and electrolyte solution every three months. In water containing large amounts
of suspended solids, for example, open recirculating cooling water, membrane cleaning or
replacement will be more frequent.
7.2.2 
Clean the membrane
Clean the membrane with water sprayed from a wash bottle.
NOTICE
Do not use tissues to clean the membrane.
7.2.3 
Replace the electrolyte solution and membrane
 WARNING
Harmful substance
Fill solution may cause irritation. May be harmful if swallowed.
Read and follow the instructions.
Procedure
1. Unscrew the membrane retainer.
74
Reference Manual
00809-0100-3412
Emerson.com