Emerson Solu Comp Xmt-P-FF/FI Instruction Sheet

Fieldbus two-wire ph/orp transmitter

Advertisement

Quick Links

Instruction Sheet
PN 51A-Xmt-P-FF-FI/rev.A
January 2006
F
OUNDATION
pH/ORP Transmitter
For additional information, please refer to the Instruction Manuals CD shipped with this
product, or visit our website at www.emersonprocess.com/raihome/liquid/.
Your purchase from Rosemount Analytical, Inc. has resulted in one of the finest instruments available for your
particular application. These instruments have been designed, and tested to meet many national and interna-
tional standards. Experience indicates that its performance is directly related to the quality of the installation
and knowledge of the user in operating and maintaining the instrument. To ensure their continued operation to
the design specifications, personnel should read this manual thoroughly before proceeding with installation,
commissioning, operation, and maintenance of this instrument. If this equipment is used in a manner not speci-
fied by the manufacturer, the protection provided by it against hazards may be impaired.
• Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life;
personal injury; property damage; damage to this instrument; and warranty invalidation.
• Ensure that you have received the correct model and options from your purchase order. Verify that this man-
ual covers your model and options. If not, call 1-800-854-8257 or 949-757-8500 to request correct manual.
• For clarification of instructions, contact your Rosemount representative.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Use only qualified personnel to install, operate, update, program and maintain the product.
• Educate your personnel in the proper installation, operation, and maintenance of the product.
• Install equipment as specified in the Installation section of this manual. Follow appropriate local and national
codes. Only connect the product to electrical and pressure sources specified in this manual.
• Use only factory documented components for repair. Tampering or unauthorized substitution of parts and
procedures can affect the performance and cause unsafe operation of your process.
• All equipment doors must be closed and protective covers must be in place unless qualified personnel are
performing maintenance.
• If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against
hazards may be impaired.
®
Fieldbus Two-Wire
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Model Solu Comp
®
Xmt-P-FF/FI

Advertisement

loading

Summary of Contents for Emerson Solu Comp Xmt-P-FF/FI

  • Page 1 Instruction Sheet PN 51A-Xmt-P-FF-FI/rev.A ® Model Solu Comp Xmt-P-FF/FI January 2006 ® Fieldbus Two-Wire OUNDATION pH/ORP Transmitter For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit our website at www.emersonprocess.com/raihome/liquid/. ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Your purchase from Rosemount Analytical, Inc.
  • Page 2: Specifications - General

    MODEL Xmt-P SPECIFICATIONS SPECIFICATIONS - GENERAL A manual two-point calibration is made by immersing the sensor in two different buffer solutions and entering the pH Case: ABS. panel mount, polycarbonate (pipe/surface values. The microprocessor automatically calculates the mount). Both versions are NEMA 4X/CSA 4 (IP65). slope which is used for self-diagnostics.
  • Page 3 MODEL Xmt-P SPECIFICATIONS SENSOR COMPATIBILITY CHART pH/ORP SENSOR DIAGNOSTIC CAPABILITY 320HP-58 Glass only 328A Glass only Glass only Glass only Glass only 381 pHE-31-41-52 Glass only 381+ Glass and Reference 385+ Glass and Reference 389-02-54 / 389VP-54 Glass only 396-54-62 / 396VP Glass only 396P-55 / 396PVP-55 Glass and Reference...
  • Page 4 QUICK START GUIDE FOR MODEL SOLU COMP Xmt-P-FF/FI TRANSMITTER 1. Refer to page 5 for installation instructions. 2. Wire pH or ORP sensor to the transmitter. See Figure 6 for panel mount; Figure 7 for pipe or surface mount. Refer to the sensor instruction sheet for details.
  • Page 5: Unpacking And Inspection

    Save the box. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Emerson Process Management immediately. INSTALLATION 1. Although the transmitter is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme FIGURE 1.
  • Page 6 MODEL Xmt-P INSTALLATION Panel Mounting. MILLIMETER INCH FIGURE 3. Panel Mount Installation Access to the wiring terminals is through the rear cover. Four screws hold the cover in place.
  • Page 7 MODEL Xmt-P INSTALLATION Pipe Mounting. MILLIMETER INCH FIGURE 4. Pipe Mount Installation The front panel is hinged at the bottom. The panel swings down for access to the wiring terminals.
  • Page 8 MODEL Xmt-P INSTALLATION Surface Mounting. MILLIMETER INCH FIGURE 5. Surface Mount Installation The front panel is hinged at the bottom. The panel swings down for access to the wiring terminals.
  • Page 9 MODEL Xmt-P INSTALLATION FIGURE 6. Loop Power and Sensor Wiring - Panel Mount...
  • Page 10 FIGURE 7. Loop Power and Sensor Wiring - Pipe/Surface Mount...
  • Page 12 ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE right now. http://www.raihome.com Specifications subject to change without notice. Credit Cards for U.S. Purchases Only. Emerson Process Management Liquid Division 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250 http://www.raihome.com...