8. Choose
9. Follow the screen prompts. Once the reading is stable, press ENTER. Take the
10. Determine the free chlorine concentration in the sample and enter the value in the
11. To return to the main display, press MENU and then EXIT.
Rosemount FCL 1056
In Process Cal
sample and press ENTER.
At this point, the transmitter stores the present sensor current and temperature and
uses those values in calibration.
screen below.
See
Purpose of chlorine calibration
If the calibration was successful, the live reading changes to the value entered in
Step 10, and the display returns to the screen in Step 6. If the sensitivity is too far
outside the range of expected values the following screen appears.
The transmitter doesn't update the calibration. For troubleshooting assistance, see
Troubleshooting when no error message is showing -
.
for sampling and testing precautions.
Calibration
October 2023
chlorine.
63