2. Remove the sensor from the flow cell. If the process and buffer temperatures are
3. Press MENU.
4. Press ENTER.
5. Choose the sensor you wish to calibrate.
6. Choose
7. Choose
8. Choose
9. Choose
10. Rinse the sensor with water and place it in the first buffer. Be sure the glass bulb
11. Press ENTER.
12. If the displayed value is not correct, press Up or Down until the correct value is
13. Press ENTER.
14. If the displayed value is not correct, press Up or Down until the correct value is
15. Press ENTER.
16. To make a change, move the cursor to the desired line and press ENTER.
17. Press ENTER to store the change.
18. To return to the main display, press MENU and then EXIT.
5.4.4 
Calibrate pH sensor manually
Procedure
1. Obtain two buffer solutions.
Rosemount FCL 1056
appreciably different, place the sensor in a container of tap water at the buffer
temperature.
Do not start the calibration until the sensor has reached the buffer temperature.
The main Menu screen appears. The cursor is on
is the chlorine sensor.
Sensor 1
.
pH
.
Buffer Cal
.
Auto
Start Auto Cal
If you wish to change the stability criteria or the pH buffer list from the default
values, choose
instead and go to Step 16. The default stability is defined as
Setup
a less than 0.02 change in 10 seconds. The default buffer list is Standard. See
5-1.
and reference junction are completely submerged. Swirl the sensor.
Once the pH reading meets the stability requirements, the screen changes to show
the nominal pH of the buffer. The nominal pH is the pH value at 77 °F (25 °C).
showing.
Once the pH reading meets the stability requirements, the screen changes to show
the nominal pH of the buffer.
showing.
If the calibration is successful, a screen is displayed for five seconds. The display
then returns to the screen in Step 7. If the calibration is not successful, the existing
calibration data is not changed. A screen appears identifying the error (high slope,
low slope, or offset error). For troubleshooting, see
message is showing -
pH. If you chose setup in Step 9, the screen below appears.
A screen appears in which the present setting can be edited.
Ideally, the buffer pH values should bracket the range of pH to be measured.
Calibrate
is the pH sensor.
Sensor 2
.
Calibration
October 2023
.
Troubleshooting when no error
Table
67