If interference due to patient activity presents a problem,
try one or more of the following to correct the problem:
Verify that the OxiMax sensor is properly and
securely applied.
Move the OxiMax sensor to another site.
Use an adhesive to the OxiMax sensor.
Use a new the OxiMax sensor with fresh adhesive
backing.
Keep the patient still, if possible.
If interference due to poor perfusion presents a problem,
consider using the MAX-R OXIMAX sensor or the
MAXFAST OXIMAX sensor. The MAX-R OXIMAX
sensor obtains measurements from the nasal septal
anterior ethmoid artery, an artery supplied by the internal
carotid.
These
measurements when peripheral perfusion is relatively
poor.
OXIMAX
- 84 -
Performance Considerations
sensors
may
obtain