H100N Pulse Oximeter User Manual
connector should not be dipped.
7
After monitoring, disinfect the probe according
to the instruction described in the user manual.
8
The materials with which the patient or any
other person can come into contact conform with
the standard of ISO10993.
9
The pictures and interfaces in this manual are
for reference only.
10 A functional tester cannot be used to assess
the accuracy of the pulse oximeter probe or the
pulse oximeter monitor.
11 If there is independent demonstration that the
particular calibration curve is accurate for the
combination of a pulse oximeter monitor and a
pulse oximeter probe, then a functional tester can
measure the contribution of a monitor to the total
error of a monitor/probe system. The functional
tester can then measure how accurately a particular
- 16 -
Safety Information