pigmented cream.
Failure to cover the OxiMax sensor site with
opaque materials in high ambient light conditions.
Venous pulsation.
Dysfunctional hemoglobin.
Low perfusion.
Loss-of-pulse signal occurs for the following reasons:
The OxiMax sensor is applied too tightly.
Defibrillation.
A blood pressure cuff is inflated on the same
extremity as the one with the OxiMax sensor attached.
There is arterial occlusion proximal to the OxiMax
sensor.
Poor peripheral perfusion.
Loss of pulse/cardiac arrest.
To use the OxiMax sensor:
Select an appropriate OxiMax sensor.
Performance Considerations