2 Safety and disposal
2 Safety and disposal
Please observe the Testo information document (enclosed with the product).
3 Product-specific safety instructions
The condensate may be acidic.
Risk of burns to the hands!
- Wear acid-resistant safety gloves, glasses and overalls to empty the
condensate.
Make sure that the condensate has been fully emptied out of the condensate
•
trap before the measuring instrument is stored for a long time.
•
Before disposing of the product, the condensate trap must be emptied and
the condensate in the crude gas tube disposed of in a suitable container.
When testing a gas pipe, pay attention to the following:
•
Dangerous mixture of gases
Danger of explosion!
- Make sure there are no leaks between the sampling point and the measuring
instrument.
- Do not smoke or use naked flames during the measurement.
Acid in the sensors.
May cause chemical burns.
- Do not open the sensors.
Eye contact: Rinse the affected eye thoroughly under running water for 10
minutes, keeping the eyelids wide open and protecting the unaffected eye.
Remove contact lenses wherever possible.
8
CAUTION
WARNING
CAUTION