Settings for Input Signal
This section describes the settings for controlling by using external input signals.
Settings for the Active Direction of an Input Signal
Set ON or OFF for an external input signal to an active state for each of the input wires.
FUN mode-[I/O SET]-[INPUT]-[ACTIVE]
Setting
IN0 to IN3
OFF
ON
Changing the Assignment of Input Signals
You can select the function to be assigned to external input signals IN0 to IN3 from two
patterns.
• If you use SmartMonitor ZS, you can change the function assignments of IN2 and IN3 if [BANK] is
selected. For details, refer to the Help for SmartMonitor ZS.
• These settings are stored in each bank and must be set separately for each bank.
FUN mode-[I/O SET]-[INPUT]-[MODE]
Setting
NORMAL
BANK
• Combination of Banks
With the combination of bank inputs A and B, you can select any banks.
Bank input A
OFF
OFF
ON
ON
• Bank switching is begun 0.5 seconds after the input state changes.
• At most it takes about 30 seconds to switch banks.
• During bank switching the BUSY output becomes ON.
• If the bank mode is set to [JUDGMENT VALUE], the bank cannot be switched at the external signal
input because the number of banks increases to 32.
The input wire is regarded as active when it is OFF.
The input wire is regarded as active when it is ON. (default value)
Select this to use external input function as in standard applications so far. (default
value)
IN0
IN1
External trigger
Reset input
(timing) input
Select this to switch banks using external inputs.
IN0
IN1
Bank input A
Bank input B
Bank input B
OFF
ON
OFF
ON
Settings for Input Signal
Description
Description
IN2
IN3
LD-OFF input
Zero reset input
IN2
IN3
LD-OFF input
Zero reset input
Bank to be Selected
BANK1
BANK2
BANK3
BANK4