RF Vector Network Analyzer
N9923AN
8. Position the rear case directly over, and aligned with, the front case and the side panels
as shown.
9. Make sure the battery cable is tucked in between the measurement group assemblies
and the blank side panel and that it is not in the path of the screw hole indicated.
10. Firmly press the cases together being careful not to damage the gaskets.
11. Install six new screws (item 7) to secure the rear case to the front case. These new
screws are provided in the
12. Tighten the six screws (item 8) that secure the RF top panel.
Figure 7-13 Combining the Front and Rear Cases (3)
13. Reinstall the battery. Refer to
Service Guide N9923-90005
Repair and Re-assembly
"Removing and Replacing the Main Battery" on page
Repair and Replacement Procedures
Removing and Replacing the Case Group Parts
Kit.
7-5.
7-17