Agilent Technologies FieldFox N9923AN Service Manual page 102

Rf vector network analyzer
Table of Contents

Advertisement

Repair and Replacement Procedures
Removing and Replacing the Front Panel Group Assemblies
Replacing the Real Time Clock (RTC) Battery Located on the A3 Front
Panel Interface Board (FPIB)
The RTC battery is a Lithium-ion battery and must be disposed of properly.
IMPORTANT
Refer to
disposing of this battery .
A soldering tool is required for this procedure.
NOTE
1. Perform the steps under
2. Unsolder the RTC battery from the A3 FPIB.
3. Insert and solder the new RTC battery into place.
4. Perform the steps under
Figure 7-17 Replacing the RTC Battery
Post-replacement Procedure
1. Reinstall the measurement group assemblies (RF top panel, A4 RF board, A5 system
board, and A6 SOM board) as one unit with the I/O panel attached. Refer to
and Replacing the Measurement Group Assemblies" on page
2. Combine the front and rear cases. Refer to
page
7-15.
7-24
"Battery Disposal" on page 1-7
"Pre-replacement Procedure" on page
"Post-replacement Procedure" on page
for important instructions on
7-18.
7-25.
"Combining the Front and Rear Cases" on
Service Guide N9923-90005
RF Vector Network Analyzer
N9923AN
7-24.