Agilent Technologies FieldFox N9923AN Service Manual page 89

Rf vector network analyzer
Table of Contents

Advertisement

RF Vector Network Analyzer
N9923AN
Replacing the Battery Cable
1. Separate the front and rear cases. See
page
7-8.
2. Note the orientation of the cable for installation of the new cable, then, using a small
flat-blade screwdriver to move the cable connector, release the retention clips on the
cable connector and remove the cable from the rear case. Do not over bend the retention
clips as they could break.
Figure 7-7
Replacing the Battery Cable
3. When installing the new battery cable, make sure that the connector snaps into place.
4. Reinstall the rear case onto the front case. See
Replacing the Rear Case
1. Separate the front and rear cases. See
2. Carefully remove the serial number label and the Windows CE label from the back of
the old rear case and apply the labels to the new rear case.
3. Remove the kickstand assembly from the old rear case and install it on the new rear
case. See
Replacing the Kickstand, Hinge Cover, and
4. Remove the battery cable from the old case and install it in the new case. See
the Battery
Cable.
5. Install the new rear case onto the front case. See
Replacing the Strap Handle
1. Separate the front and rear cases to the positions shown in
the Front and Rear
2. Note the orientation of the strap handle, then lift it off of the dowel pins at each end.
Service Guide N9923-90005
"Separating the Front and Rear Cases" on
Separating the Front and Rear
Cases.
Repair and Replacement Procedures
Removing and Replacing the Case Group Parts
Combining the Front and Rear
Hinge.
Combining the Front and Rear
Figure
7-5. See
Cases.
Cases.
Replacing
Cases.