Agilent Technologies FieldFox N9923AN Service Manual page 113

Rf vector network analyzer
Table of Contents

Advertisement

RF Vector Network Analyzer
N9923AN
Reinstalling All Measurement Group Assemblies as One Unit
1. Reinstall the I/O side panel, if needed. Refer to
and Gasket" on page
2. Position the measurement group assemblies to the side of the front case as shown in
Figure 7-18 on page
One Unit" on page
3. Reconnect the LCD to system board cable at the system board connector and place a
strip of adhesive tape over the connector to secure it. This strip of adhesive tape is
provided in the
Repair and Re-assembly
LCD, the LCD Shield, and LCD Cables" on page
4. Position the measurement group assemblies on top of the front case and align the I/O
side panel in the groove in the front case being careful not to damage the gasket.
5. Position (tuck) the LCD to system board cable in between the LCD shield and the A5
system board so that it will not interfere with the cases when they are combined.
6. Reconnect the system board to front panel interface board cable at the front panel
interface board connector as shown.
Service Guide N9923-90005
7-12.
7-26. Refer to
"Removing All Measurement Group Assemblies as
7-25.
Repair and Replacement Procedures
Removing and Replacing the Measurement Group Assemblies
"Replacing the I/O Side Panel, Doors,
Kit. If necessary, refer to
7-20.
"Replacing the A2
7-35