10.Reinstall the rear case onto the front case. Refer to
Cases" on page
7-15.
Combining the Front and Rear Cases
1. Install a new gasket on the blank side panel being careful not to stretch it or to damage
it in any way. Also, make sure there is no slack in the gasket. A new gasket is provided
in the
Repair and Re-assembly Kit
gaskets provided). If an old gasket must be removed, carefully remove it with tweezers
being careful not to damage the groove.
2. If a new gasket has not been installed on the I/O side panel, install it at this time. It will
be necessary to remove the I/O side panel from the A5 system board. Refer to
the I/O Side Panel, Doors, and Gasket" on page 7-12
3. Position the front and rear cases side-by-side as shown.
4. Plug the battery cable, W10, into the A5 system board as indicated.
Figure 7-11 Combining the Front and Rear Cases (1)
Service Guide N9923-90005
Removing and Replacing the Case Group Parts
"Combining the Front and Rear
(the blank side panel uses the larger of the two
for instructions.
Repair and Replacement Procedures
"Replacing
7-15