RMP_OUT_00_XXX.XX_XXXXX_XXX.
XX_X
RMP_OUT_02_XXX
Note:
 For "_" " " (space character) is also admissible.
 Response from thermostat "OK" or with an error " ERR_X" (RS 485 interface e.g. "A015_OK" or with an
error "A015_ERR_X".)
 The command from the computer must be terminated with a CR, CRLF or LFCR.
 The response from the thermostat is always terminated with a CRLF.
 After each command sent to the thermostat, it is necessary to wait for the reply before sending another
command. This ensures that the sequencing of inquiries and answers is clear.
CR = Carriage Return (Hex: 0D); LF = Line Feed (Hex: 0A)
Admissible data formats:
-XXXX.XX
-XXXX.X
-XXX.XX
-XXX.X
-XX.XX
-XX.X
-X.XX
-X.X
-.XX
-.X
19.7.2

Interface read commands

Command
IN_PV_00
IN_PV_01
IN_PV_03
IN_PV_04
IN_PV_10
IN_PV_13
IN_SP_00
IN_SP_01
IN_SP_02
IN_SP_03
IN_SP_04
IN_SP_05
IN_PAR_00
09/2021
Sets programmer segment (temperature, time, tolerance, and pump level). A segment is appended
and assigned appropriate values.
Number of program loops: 0 = endless / 1 – 250.
-XXXX.
-XXXX
-XXX.
-XXX
-XX.
-XX
-X.
-X
.XX
.X
Query of the bath temperature (outflow temperature)
Query of the controlled temperature (int./ext., Pt/ext., Analog/ ext. serial).
Query the external temperature TE (Pt100).
Query the external temperature TE (Analog Input).
Query of the bath temperature in 0.001 °C.
Query the external temperature TE (Pt100) in 0.0001 °C.
Query of the temperature set value.
Query of the pump power level.
Query of cooling mode (0 = OFF / 1 = ON / 2 = AUTOMATIC).
Query of the overtemperature switch-off point.
Query of the outflow temperature limit TiH.
Query of the outflow temperature limit TiL.
Query of the control parameter Xp.
ECO Gold
XXXX.XX
XXXX.X
XXX.XX
XXX.X
XX.XX
XX.X
X.XX
X.X
Meaning
XXXX.
XXXX
XXX.
XXX
XX.
XX
X.
X