Horiba Scientific LAQUAtwin B-74 Series Manual page 40

Table of Contents

Advertisement

Quick Operation Guide
This quick operation guide introduces the basic operations.
Refer to the respective chapters for further information.
Turn ON
① 
Press the ON/OFF button over 2 seconds to turn ON the meter.
Calibration
② 
①  Pour the 2000 ppm standard solution. ②  Close the light shield cover.
③   Press the CAL button over 2 seconds.
When
   and
④  Clean the sensor with water.
Perform calibration at least once a day
for accurate measurement.
light up, the calibration is completed.
⑤  Press the MEAS button to enter
the measurement mode.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Laquatwin b-741Laquatwin b-742Laquatwin b-743