■ DATA WINDOW
This item is for setting the change in the state of the data (measured value) as the log-
ging start trigger. Set the upper/lower limit value. When the measured value falls inside
or outside of that range, this is taken to be the start trigger.
FUN Mode-[LOGGING]-[TRIGGER]-[START]-[WINDOW]
Setting
SOURCE
UPPER
LOWER
MODE
DELAY
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Setting the Logging Conditions
Select the source to be used as the logging start trigger.
Range: SourceA to J
Sets the upper limit value of the specified range.
Range: -999.999 to 999.999
Sets the lower limit value of the specified range.
Range: -999.999 to 999.999
Select either of the measured value inside or outside the specified range
is to be used as the start trigger.
Range: IN, OUT (default: IN)
At IN
Logging start
At OUT
Logging start
Set the delay time when you want to start logging a little before or a little
after the start trigger.
Range: -9999 to 9999 (ms) (default value: 0 ms)
-9999ms
Delay time
Start trigger
At -9999ms
Logging start
At 9999ms
Section3
Description
Measured value
Specified range
+9999ms
Measured value
Specified range
Logging start
User's Manual
ZS-DSU
3-19