Section3
Setting the Logging Conditions
■ DATA SLOPE
This item is for setting the change in the state of the data (measured value) as the log-
ging start trigger. Set the trigger level. When the data falls above or below this preset
trigger level value, this is taken to be the start trigger.
FUN Mode-[LOGGING]-[TRIGGER]-[START]-[DATA]
SOURCE
EDGE
DELAY
ZS-DSU
3-18
User's Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Setting
Select the source to be used as the logging start trigger.
Range: SourceA to J
Select whether above or below the trigger level is to be used as the start
trigger. Then, also set the trigger level.
Range: UP, DOWN (default: UP)
At up edge
At down edge
Set the delay time when you want to start logging a little before or a little
after the start trigger.
Range: -9999 to 9999 (ms) (default value: 0 ms)
At -9999ms
At 9999ms
Description
Logging start
Logging start
-9999ms
+9999ms
Delay time
Start trigger
Logging start
Logging start
Trigger level
Measured value
Trigger level
Measured value