Pattern Trigger
The pattern trigger identifies condition by looking for a specified pattern. The pattern trigger can be
expanded to incorporate delays similar to other triggers. Pattern durations are evaluated using a timer.
The timer starts on the last edge that makes the pattern "true".
Potential triggers occur on the first edge that makes the pattern false, provided that the time qualifier
criterion has been met. The oscilloscope provides 4 patterns: logical AND, OR, NAND and NOR combination
of the channels. Each channel can set to low, high or invalid.
Configure Pattern Trigger
1. Press the
button to enter the trigger menu.
2. Press the
button to select Level Type.
A dropdown menu containing the available options will appear.
3. Use the
Universal Knob
Repeately pressing
4. Press the
softkey to select Next Page
Figure 5.25 Pattern Trigger
Figure 5.26 Level Type
to navigate the available options.
will navigate between the available choices.
Trigger