29. Measure
The measured values can be detected on each channel simultaneously. Only if the waveform channel is in the
ON state, the measurement can be performed. The automatic measurement cannot be performed in the
following situation: 1) On the saved waveform. 2) On the mathematical waveform. 3) On the XY format. 4) On
the Scan format.
Measure the frequency, the peak-to-peak voltage of the Channel CH1 and the mean, the RMS of the Channel
CH2, following below steps:
1. Press the Measure button to show the automatic measurement function menu.
2. Press the H1 button to display the Add menu.
3. Press the F2 button and choose CH1 as the source.
4. Press the F1 button, the type items will display at the left of screen, and turn the M knob to choose
Period.
5. Press the F4 button, the period options added completes.
6. Press the F1 button again, the type items will display at the left of screen, and turn the M to choose Freq.
7. Press the F4 button, the frequency added completes, finish setting of CH1.
8. Press the F2 button and choose CH2 as the source.
9. Press the F1 button, the type items will display at the left of screen, and turn the M to choose Mean.
10. Press the F4 button, the Mean added completes.
11. Press the F1 button, the type items will display at the left of screen, and turn the M to choose PK-PK.
12. Press the F4 button, the PK-PK added completes, finish setting of CH2.
The measured value will be displayed at the bottom left of the screen automatically (see Fig. 68).
Fig. 68 automatic measurement
-152-