Digital Module Operation 7-18
DIG:SYNC:POINTS (@7(3)),256
Outputs 44 to port 4, 55 to port 5, 88 to port 8, and 99 to port 9:
DIG:OUTPUT (@7(4,5,8,9)),44,55,88,99
Commands the 1260-114 to read the data. The reply will be a set
of 4 data byte values, in decimal format, separated by commas:
DIG:INPUT? (@7(6,7,10,11))
Enables the synchronous mode for module #7:
DIG:SYNC:STATE (@7),ON
After this operation is complete, generate 256 clock pulses on the
CLKIN input line of module #7.
Reads whether module #7 is still armed. If the reply is "1", then the
module is still armed, expecting more CLKIN pulses. If the reply is
"0", then the module is disarmed:
DIG:SYNC:STATE? (@7)
Read the data bytes captured by port 2 during the synchronous
test. The reply will be a set of 5 data bytes, in decimal format,
separated by commas:
DIG:SYNC:DATA? (@7(2))
Read the data bytes captured by port 3 during the synchronous
test. The reply will be a set of 256 data bytes, in decimal format,
separated by commas:
DIG:SYNC:DATA? (@7(3))
1256E User Manual
Racal Instruments © 2002