Front-Panel Controls
PROBLEM
Display goes blank
(was working).
No response to front-panel
controls.
Appendix C-2
POSSIBLE CAUSES
Power has been
interrupted.
1256 screen saver has
been activated.
Key is not currently
enabled. Enabled keys
have their LEDs
illuminated.
1256 is in Automatic Local
lockout mode.
1256 is in Manual Lockout
mode.
1256 User Manual
WHAT TO DO
Verify that the 1256 AC input is
connected to a live mains
outlet.
Verify that the 1256 power
switch is still in the ON
position.
Screen saver activation is
normal after a period of no
front-panel control activity. To
restore the display, press any
key on the front panel.
To increase the delay time, or
disable the screen saver, refer
to Chapter 3: USING THE
FRONT-PANEL CONTROLS,
under the Settings heading.
Note that leaving the screen
saver enabled maximizes the
life of the fluorescent display.
Read display options carefully.
Not all keys are valid for all
menus.
If display reads:
*REMOTE OPERATION*
(RS-232 is active)
<Local Mode
then press key 4 (bottom) to
restore display and controls.
If display reads:
*REMOTE OPERATION*
(RS-232 is active)
Local Mode Lockout
then the 1256 is in Manual
Lockout mode. To restore
controls, send a
SYST:KLOCK OFF command
to the 1256 via remote
interface.