PROBLEM
Text from 1256 to remote
terminal overwrites itself,
staying on one line.
Text from 1256 to remote
terminal is double-spaced
(lines of text are separated by
blank lines).
Remote terminal does not
display characters that are
typed.
Appendix C-4
POSSIBLE CAUSES
Baud rate too high for cable
and terminal being used.
Cable is excessively long,
or not of high quality.
Termination characters of
1256 not compatible with
terminal setup.
Termination characters of
1256 not compatible with
terminal setup.
Echo mode of remote
terminal is not set to half-
duplex (local echo).
HyperTerminal (included
with Microsoft Windows)
can fail to echo characters
properly.
1256 User Manual
WHAT TO DO
Change to a lower baud rate.
Make sure that the new 1256
baud rate matches the new
terminal baud rate (to set the
1256 baud rate, see Chapter
4, GPIB and RS-232
Interfaces, under the heading
Setting the Baud Rate).
Use a high-quality, shielded
cable.
Set 1256 to use carriage return
and line feed (CR-LF) as
termination characters (see
Chapter 4, GPIB and RS-232
Interfaces, under the heading
Setting the Message
Termination Characters).
Set 1256 to use only carriage
return (CR) as termination
character (see Chapter 4,
GPIB and RS-232 Interfaces,
under the heading Setting the
Message Termination
Characters).
Set terminal for half-duplex
(local echo) mode. Consult the
terminal user's manual for
instructions.
Download newer version,
known as "HyperTerminal
Private Edition" from:
http://www.hilgraeve.com
Racal Instruments © 2002