GENERAL SAFETY

SAFETY WARNING
This analyser extracts combustion gases that may be toxic in relativity
low concentrations. These gases are exhausted from the bottom of the
analyser. This analyser must only be used in well-ventilated locations
by trained and competent persons after due consideration of all the
potential hazards.
Portable gas detectors users should conduct "bump" tests before
relying on units to verify atmospheres are free from hazard.
A "bump" test is a way to check an instrument works within acceptable
limits by briefly exposing it to known gas mixtures to change the output
of all sensors present.
Note: This is different from a calibration where the instrument is also
exposed to known gas mixtures but allowed to settle to a steady figure
with readings adjusted to the stated gas concentration of the test gas.
Protection Against Electric Shock (In accordance with EN 61010-1:
2010):
This analyser is designed as Class III equipment and should only be
connected to SELV circuits. The battery charger is designated as:
Class II equipment
Installation category II
Pollution degree 2
Indoor use only
Altitude to 2000m
Ambient temperature 0°C-40°C
Maximum relative humidity 80% for temperatures up to 31°C
decreasing linearly to 50%RH at 40°C
Mains supply fluctuations not to exceed 10% of the nominal
voltage
10 KANE 458s MANUAL