GENERAL SAFETY

SAFETY WARNING
Your analyser extracts combustion gases that may be toxic in relativity low
concentrations.
These gases are exhausted from the back and bottom of the analyser.
This analyser must only be used in well-ventilated locations by trained and
competent persons after due consideration of all potential hazards.
Portable gas detectors should conduct "bump" tests before relying on units to
verify atmospheres are free from hazards.
A "bump" test checks an analyser works within acceptable limits by briefly
exposing it to known gas mixtures to change output of sensors present.
NOTE: This is different from a calibration where your analyser is
exposed to known gas mixtures but allowed to settle to a steady figure
with readings adjusted to the test gas concentration.
Protection Against Electric Shock - (In accordance with EN 61010-1:2010).
This analyser is designed as Class III equipment and should only be
connected to SELV (Safety Extra Low Voltage) circuits.
The battery charger is designated as:
Class II equipment
Installation category II
Pollution degree 2
Indoor use only
Altitude to 2000m
Ambient temperature
Maximum relative humidity 80% for temperatures up to 31°C
decreasing linearly to 50%RH at 40°C
Mains supply fluctuations not to exceed 10% of the nominal voltage
0°C -40°C