6.4.21
Heated Circuit Fault Alarm
Property
Alarm text
Alarm condition
Priority
Ventilator action
Reset
6.4.22 Internal Function Failure
Property
Alarm text
Priority
Alarm condition
Reset criteria
Ventilator action
Action to take
104
Alarms
User Manual
Description
Heated Circuit Fault
A Heated Circuit Fault alarm will be given if a fault in the heated
circuit electronics or temperature sensor is detected.
Medium
The ventilator will turn off the heated circuit and continue treat-
ment with the same settings.
The heated circuit must be restarted manually when the cause
of the alarm is resolved.
The alarm is dismissed when the heated circuit setting is
changed to OFF, or the treatment is stopped.
The power to the heated circuit is re-enabled when all enabling
conditions are satisfied.
Description
Int. Function Failure
High
Failure of internal function that prevents treatment or normal
operation of the ventilator.
The error code that follows the alarm text indicates the kind of
function failure.
All Internal Function Failure alarm error codes are defined and
explained in the ventilator Service Manual.
Restart the ventilator.
The ventilator will stop the treatment and shut down.
Restart the Nippy 4.
If the alarm persists or reoccurs: Take a note of the error code
and contact your supplier of the Nippy 4 .
Doc. 006553 M-6