Breas NIPPY Quick Reference Manual page 2

Advertisement

Quick Reference Guide
This is not a complete guide for the Nippy 4. See your
manual for complete instructions.
Switch On/Off
Press the On/Off button on the
side panel.
Connect the Patient Circuit
Always perform a pre-use test when the patient circuit is
replaced or modified. If the pre-use test is not
prompted at startup, you can start it from the menu.
The leakage from the mask leakage port should be at least 12
l/min at 4 cm H2O to prevent rebreathing of exhaled air.
1.
Connect the patient circuit to
the patient air outlet on the
ventilator.
2.
Connect the other end of the
patient circuit to a leakage
port or to a patient interface
with integrated leakage port.
Start/Stop Treatment
1.
To start treatment and enter
operating mode first press
and hold the Start/Stop but-
ton on the front panel.
2.
Release the Start/Stop but-
ton when the progress bar is
filled.
3.
To stop treatment and enter standby mode first press
and hold the Start/Stop button on the front panel.
4.
Release the Start/Stop but-
ton when the progress bar is
filled.
5.
Press the Audio Pause button
to stop the treatment.
i-Button
Use the i-button to get
explanations about settings and
alarms.
Connectors
Make sure that the marking is
facing up.
Pull the connector sleeve, not the
cable itself to release the
connector.
Connect the Click-in Battery (Accessory)
1.
Release the side cover by
pressing the button under
the handle.
2.
Open and remove the side
panel.
3.
Insert the click-in battery.
Make sure the latch at the
bottom of the click-in com-
partment is engaged.
4.
Close the side panel. Make
sure there is a clicking sound
to secure the side panel.