N O T E
N O T E
30
If the E4982A was turned off without proper shutdown process, Microsoft
Scandisk runs while the windows boot screens are displayed. If a serious
problem is found during the scan, execute the system recovery. For details
on executing the system recovery, refer to the Installation Guide. If the
operating system still does not boot up properly after reinstallation,
replace the HDD.
The operating system automatically checks the device drivers which are
necessary for use in the E4982A and are installed in the system before the
E4982A is shipped from Keysight factory. If the operating system does not
detect them, a message box is displayed. In this case, install the device
driver.
Step 4.
System Initialization Screen.
The system initialization screen is displayed as shown in
Figure
11. This screen shows the status of the items
initialized at start up.
Figure 11 System Initialization Screen
If the display whites out, is entirely blue or a dialog box
appears, a mass storage problem is suspected. Try to
perform the mass storage recovery procedure.