Troubleshooting
N O T E
N O T E
36
To perform the diagnostic test properly, the following conditions must be
met:
Environmental temperature:
Warm up time: > 30 minutes
Do not operate front panel keys, keyboard and mouse during the
diagnostic test. Changing the instrument setting while the diagnostic test
is in progress will cause incorrect test results.
Step 1.
Press
Step 2.
Click Service - External Test.
The external test warning screen as shown in
displayed. Check the warning conditions and click Start
Test.
Figure 17 External Test 1/4
Step 3.
Select Execute All or Execute Single.
23°C ±5°C
key.
Figure 17
is
LCR Meter