Keysight Technologies MY53280000 Installation Notes page 23

Mxa signal analyzer
Table of Contents

Advertisement

7. Refer to
Figure
14. Pull the AIF/DIF assembly part way out of the chassis. Disconnect cable
W15 from the analyzer.
8. Completely remove W15 from the analyzer.
Figure 14
AIF/DIF Assembly Partial Removal
9. Locate W36 in the Opt EXM/HL6 Cable Kit. This cable will have part number 8121-1862 and
will be labeled "902" on one end and "7" on the other. Connect the end labeled "902" to
A15J902 on the A15 EFEC Assembly. Connect the other end to A13J7 on the A13 Front End
Assembly. Torque the connector on A13J7 to 10 inch-pounds.
10.Locate W37 in the Opt EXM/HL6 Cable Kit. This cable will have part number 8121-1401 and
will be labeled "900" on one end and "100" on the other. Connect the end labeled "900" to
A15J900 on the A15 EFEC Assembly. Route the cable along the left side of the instrument
above the fans. Route the cable down behind the rear of the second (rearmost) fan. Connect
the cable to A2J100 on the A2 EAIF Assembly
11.Slide the AIF/DIF assembly into the slot at the rear of the instrument and push on the
assembly to mate the connectors to the motherboard assembly.
Installation Note N9020-90230
Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328
23