Verify Hardware Installation

1. Verify the spectrum analyzer application loads and sweeps as expected.
2. Press System, Show Hard ware on the analyzer and verify that the Wideband Analog IF and
Wideband Digital IF boards identify themselves as:
— WB Analog IF (Hw Id = 38)
— WB Digital IF (Hw Id = 40)
Verify Optional Functionality
1. Press MODE/MEAS, I/Q Analyzer (Basic), OK.
2. Press MEAS SETUP, IF Path, and select the IF Path drop-down menu.
3. Verify that selections appear labeled either "85 MHz", "125 MHz" or "160 MHz" appear,
depending upon whether upgrade Option BU1, BU2, or BU3, respectively, was installed.
4. Select IF Path of either 85 MHz, 125 MHz, or 160 MHz, as appropriate.
5. Press FREQ, Center Frequency, 4.8 GHz. If the frequency range option is 503, set the Center
Frequency to 1 GHz.
6. Press Input/Output, RF Calibrator, and select 4.8 GHz from the RF Calibrator drop-down
menu. If the frequency range option is 503, apply a 1 GHz, -28 dBm CW signal from a signal
generator to the analyzer input.
7. Press FREQ, Span, 85 MHz.
8. The analyzer should display a signal in the center of the screen with an amplitude of
approximately -28 dBm (see
Installation Note N9020-90254
Figure
25).
Installation Procedure