Keysight Technologies MY53280000 Installation Notes page 25

Mxa signal analyzer
Table of Contents

Advertisement

Add Cables to A13 Front End and A15 Front End Controller
1. Locate the flexible coax assembly in the Opt EXM/HL6 Cable Kit with Markers that is labeled
"8120-2027". The ends should be labeled "903" and "13". This is W26.
2. Connect the end of W26 that is labeled "903" to A15J903.
3. Refer to
Figure
16. Connect the end of W26 that is labeled "13" to A13A1J13. J13 is one of the
connectors along the right side of A13, near the A12 BYFA (if present).
4. Locate semi-rigid coax cable, part number N9020-20241, in the upgrade kit. This is W28.
Connect the end with the SMA male connector A13J8, with the SMA female connector
pointing towards where the front panel would be. The long, straight section of W28 should be
parallel to the casting of the A13 Front End and parallel to the side chassis. Torque the cable
nut to 10 inch-pounds.
Figure 16
Orientation of W28 in MXA
Installation Note N9020-90230