2. Refer to
Figure
7. Pull the front frame carefully away from the chassis. Remove the ribbon
cable W1 from the A8 Motherboard.
W1 may have locking springs on each side. Depress the spring on each side of the connector to
remove from the motherboard.
Figure 7
Front Panel Cable
3. If the instrument has Option BBA (BBIQ inputs), there will be an additional cable (W24) that
will need to be removed. W24 connects to the A18 BBIQ Interface Board. Pull the Front Frame
Assembly carefully away from the chassis. Remove the ribbon cable W1 from the
motherboard.
Installation Note N9020-90230
Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328
15