Adjustments Required - Keysight Technologies MY53280000 Installation Notes

Mxa signal analyzer
Table of Contents

Advertisement

Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328
Utilities, Adjustments, and Performance Verification Tests
Calibration Software and specified test equipment is required to perform the adjustments, and can
be used to automate the performance verification testing.
Obtain Keysight X-Series Signal Analyzer Calibration Application SW, N7814A TME Calibration
Application, version E.11.04 or later. Information on how to obtain this software can be found at:
http://www.keysight.com/find/calibrationsoftware
The following tests are required to assure the installation was performed correctly. The instrument
may not have been in spec before the upgrade was begun. Performing only these tests does not
guarantee that the analyzer meets all specifications.
Utilities Required
None

Adjustments Required

— Perform all adjustments
Performance Testing Required
— Perform all performance tests
A full calibration is required to assure the instrument meets all specifications
The end user must ultimately determine whether they want a full calibration to be performed after
the installation of this upgrade or not. If a full calibration is required, arrangements regarding the
level of calibration must be made between the end user and the calibration provider.
For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your
local Keysight office access the following URL, or if in the United States, call the following
telephone number:
http://www.keysight.com/find/assist
1-800-829-4444 (8 am - 8 pm ET, Monday - Friday)
38
Installation Note N9020-90230