USER MANUAL
A6000
AC POWER SOURCE

3.19 GPIB, RS232 Communication Setting

DME-A602ABS SERIES provide remote control operation, there are GPIB or RS232
interface for selection. Only one interface can be activated at the same time:
Operation mode
Manual (front panel)
Remote (GPIB)
Remote (RS232)
How to switch manual to remote control: Send any command to AC source directly,
AC source enter remote control mode automatically.
Two ways to Switch from remote to local control: 1. Stop remote control, press
longer than 2 seconds until "SET MODE" shown on upper left on the screen; 2. Send *CLS
command.
Switch remote control between GPIB or RS232: Back to manual operation as
demonstrated above, then use another communication interface to do the control.
Parameters setting for two interfaces:
GPIB
Address
 1~30
Example how to set GPIB Address to 20:
1. Press
47
│ www.deltaww.com
Controller
(front panel)
RS232
Odd/Even parity check (Parity)
 Even parity check (EVEN)
 Odd parity check (ODD)
 No parity check (NONE)
or
to move to GPIB ADDR
Manual
Remote
(GPIB)
Baud rate (Baud Rate)
 9600
 19200
Remote
(RS232)