Gpib, Rs232 Communication Setting - Delta A1500 Series User Manual

Table of Contents

Advertisement

User Manual
A1500
AC Power Source
3. Press
4. By pressing numerical key or rotating knob, change parameter to 1.300, press
confirm the setting

3.6.2 GPIB, RS232 Communication Setting

A1500 SERIES provide remote control operation, there are GPIB or RS232 interface for
selection. Only one interface can be activated at the same time:
Operation mode
Manual(front panel)
Remote(GPIB)
Remote(RS232)
How to switch manual to remote control:Send any command to AC source directly, AC
source enter remote control mode automatically.
Two ways to Switch from remote to local control: 1. Stop remote control, press
longer than 2 seconds until "SET MODE" shown on upper left on the screen; 2. Send
*CLS command.
Switch remote control between GPIB or RS232:Back to manual operation as
demonstrated above, then use another communication interface to do the control.
Parameters setting for two interfaces:
32
│ www.deltaww.com
Save
Save
Recall
Recall
or
to move cursor to CSIN CF
Controller
Manual
(front panel)
Remote
(GPIB)
to
Enter
Remote
(RS232)
/
Unlock

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dme-acs1152b series

Table of Contents