Technical data

Technical data

WATER SUPPLY SYSTEM
Temperature
max. 95 °F (35 °C)
Pressure
min. 29 psi - max. 124.7 psi (min. 2 bar – max. 8.6 bar)
Flow
min. 0.066 - max. 0.132 gal/min (min. 0.25 – max. 0.5 l/min)
POWER SUPPLY SYSTEM
Nominal voltage and Max.
200–240 V ac (±10%), 50/60 Hz, 10 A, single-phase
current
100–125 V ac (±10%), 50/60 Hz, 12 A, single-phase
Overvoltage category
II
Protection required
Suitable circuit breaker and a Ground Fault Circuit Interrupter
(GFCI). All protection devices must be certified according to
applicable standard. A grounded connection is essential.
Communication with other
2 USB ports - 1 LAN port (optional)
devices
Features
Fully micro-processor controlled, process evaluation system
according to EN13060.
Programmable standby mode.
Max. heat output
3000 kJ/h
INSTALLATION REQUIREMENTS
From +41 °F to +104 °F (from +5 °C to +40 °C)
Working temperature
Working relative humidity
Max. RH 80% up to 88 °F (31 °C), linearly decreasing to 50% at 104
°F (40 °C)
92
From -4 °F to +140 °F (from -20 °C to +60 °C) / 0–90 % (with empty
Storage temperature / rel.
humidity
tanks)
Max altitude
3000 m asl
8.7 psi (0.6 bar)
Min. atmospheric
pressure
Overall dimensions
W: 19.3''/H: 17.9''/D: 24.2'' (W: 49 cm/H: 46 cm/D: 62 cm)
W: 22.5''/H: 18.9''/D: 20.1'' (W: 57 cm/H: 48 cm/D: 51 cm)
Min. space required
(feet in forward position)
Size of the door
W: 21.5''/H: 16.4''/D: 15.6'' (W: 55 cm/H: 42 cm/D: 38 cm)
movement
Weight empty
105.8 lbs (48 kg)
Max. weight (fully loaded) 155 lbs (70.3 kg)
41.5 kN/m
Weight per support area
Environment pollution
Degree 2
Indoor
Usage environment
STERILIZER CHAMBER
Pressure safety valve
37.7 psi (2.6 bar)
Safety thermostats
356 °F (180 °C)
Total volume
~7.4 gal, Ø: 11''/D: 18'' ~28 l, Ø: 279.5 mm/D: 456.8 mm
Usable space *
W: 9.05"/H: 9.05"/D: 15.35" (W: 230 mm/H: 230 mm/D: 390 mm)
HEPA filter
0.3 µm
STEAM GENERATOR
Pressure safety valve
72.51 psi (5 bar)
Safety thermostats
446 °F (230 °C)
Lexa  | Instructions for Use |  MN-111 ENG Rev14 |  03/04/2023 |  © W&H Sterilization Srl