Maintenance

1200-cycle or yearly maintenance

REPLACING THE DOOR GASKET
Notice: the door gasket needs to be replaced every 1200 sterilization cycles or once a year, whichever come first. A replacement message alerts
when replacement is due. If the consumables is replaced prior to the message, you have to reset the consumable cycle counter.
Proceed as follows:
Open the chamber door.
1
Remove the used door gasket by hand.
2
Carefully clean the seal seat and the inside face of the chamber door.
3
Note: tap
to see the animated replacement procedure.