6.2.6. Total Expected Sensor Life (EXP LIFE)
The device tries to predict its overall lifetime. You can reset the factory default.
Values
To modify the value via SIMATIC PDM:
a. Open the menu Device – Maintenance and click on Remaining Sensor
Lifetime.
b. Enter the desired value (or click on Snooze to add a year to the current
value) then click on Write to accept the change.
c. Click on Read to view the effect of the modification.
Units
Allows you to set desired units using SIMATIC PDM.
Options
6.2.7. Status (MAINT STAT)
Read only. Displays the status of the Maintenance Alerts.
Options
(read only)
1.
Data for this parameter must be entered in seconds on the device.
A5E42035355
Range: 0 to 20 years
Default: 10.00 years
Hours
Days
Years
NONE
MAINT REQ
MAINT DEM
SITRANS LR460 – INSTRUCTION MANUAL
1
None
Maintenance Required
Maintenance Demanded
Page 83