The number of echo profile samples averaged to produce a measurement.
Values
3.5.4. Range
Sensor Reference Point
3.5.4.1.
The range in front of the device (referenced from the flange face) within
which any echoes will be ignored. (This is sometimes referred to as
Blanking or Dead Zone.)
Values
The default setting of 0.6 m avoids a false echo being triggered by
condensation or other material buildup. If no buildup is anticipated (for
example, with dry cement) then the minimum setting 0.35 m can be used.
3.5.4.2.
Note:
Maximum distance from the reference point, within which an echo should
be considered valid.
Values
Page 62
Shots
Range: 1 to 25
Default: 15
(flange face)
High Calibration Point
Low Calibration Point
Near Range
Range: 0 to 103 m max.
3" HORN
4" HORN
Far Range
Far Range can extend beyond the bottom of the vessel.
Range: 0 to 103 m max.
Default: The value of Low Calibration Point plus 3 m (9.8 ft)
SITRANS LR460 – INSTRUCTION MANUAL
Level
default: 0.278 m (0.9 ft)
default: 0.335 m (1.1 ft)
Near Range
Far Range
A5E42035355